Clinical effectiveness of radiofrequency ablation with a multiple-electrode switching system for liver tumors

Yunhua Tan,Hui Liu,Jia Jiang,Qiang Wang,Kuansheng Ma,Ping Bie
DOI: https://doi.org/10.3969/j.issn.1003-5591.2016.02.004
2016-01-01
Abstract:Objective To assess the effectiveness and safety of a novel radiofrequency ablation (RFA)with a multiple-electrode switching system for the treatment of liver tumors.Methods We performed a total of 178 RFAs with a multiple-electrode switching system on 176 patients from Dec. 2014 to Dec.2015.There were a total of 304 tumors treated with an average diameter of 4.72 ± 1.7 cm (range,2.2±10.4 cm).There were 134 male patients and 42 female patients with a mean age of 52.95±11.87 years old (range,22-83).Ultrasonography or contrast-enhanced CT of the liver was performed 48 h after the surgery to determine whether the ablation was complete and whether compli-cations occurred.Results For the treatment of 304 liver tumors,the complete ablation rate of multi-e-lectrode switching system was 98.68% (300/304),and the complete ablation rate with one procedure was 98.02% (298/304).The total ablation time was 2275 min,and the mean ablation time per pro-cedure was 13.04±5.39 min,and there was a 56.24% reduction relative to that of single-electrode RFA.No hospital deaths occurred in any of the enrolled 176 patients.The complication rate was 2.35%.Conclusions Multiple-electrode switching RFA is a safe and effective method for treating liver tumors.It was here found to improve the complete ablation rate with one procedure in RFA treatment of liver tumors and to shorten the RFA time,which reduces surgical risks.
What problem does this paper attempt to address?