[Research of magnesium alloys and plga degradated in vivo and micro-ct evaluate the degradation of magnesium].
Jiang Peng,Yichi Xu,Zhen Sun,Shibi Lu,Heyong Yin,X. Sui,H. Meng,Ai-yuan Wang
DOI: https://doi.org/10.7507/1002-1892.20170111
Abstract:Objective
To explore the degradation of AZ31 magnesium alloy and PLGA in the femoral condyle through weight method. Then evaluate the degradation of AZ31 magnesium alloy by Micro-CT images and data.
Methods
40 male New Zealand white rabbits (NZWRs, body weight 2.5 kg, 3 months old) were randomly divided into four groups, 10 rabbits per group. 40 AZ31 magnesium pins and 40 PLGA pins were produced and implanted in the right and left femoral condyle in 40 rabbits respectively. Degradation was monitored by weight loss of the implants before and after implantation, Micro-computed tomography (Micro-CT) images and data analysis after 4, 8, 12 and 16 weeks. Histological analyses indicated the inflammatory responses around the implants.
Results
PLGA pins showed gray low signal, which was similar with the soft tissue around. The Micro-CT images showed that in the first 4 weeks, almost no corrosion was observed on the magnesium alloy, and the boundary of the pin was continuous and smooth. While little corrosion had occurred by week 8, the surface of the pin turned fuzzy and the corners of the pins became indistinct. Then, corrosion pitting became evident at week 12 and it became more obvious and the boundary of the pin was inconclusive at week 16. The data exported from the built-in software of the Micro-CT equipment showed precise results. The pin volume fraction decreased slowly at week 4 and 8, and from week 8 to week 12, the pin volume fraction decreased significantly faster than in the early stage. In particular, it reached 86.2% at week 16 ( P<0.05).The magnesium cylinder mineral density continuously decreased during the study period, from week 12 to 16, it had a rapidly speed ( P<0.05). However, the magnesium CT image density indicating only a slight change in CT image density ( P>0.05). The ratio of the surface area and the volume of the pins increased throughout the study period, and had a high rate at week 12 and 16 ( P<0.05). With the degradation of the pins, there was more corrosion pitting on the surface, which resulted in a decrease in the radius, with statistical difference after week 8( P<0.05). The weight results showed a slowly and inconspicuously decreasing rate of degradation in the early 8 weeks of AZ31, the decrease rate of PLGA accelerated at week 8-12, and slow down again after week 12. The decrease rate of AZ31 magnesium alloy was slowly and reposefully before 12 weeks, however, during week 12-16, it turned rapidly, especially in week 16, and the weight loss more than PLGA ( P<0.05). Histological analyses indicated that there were slight inflammatory responses around the AZ31 pins in the early stage and disappeared later, and there no inflammatory responses around the PLGA implants.
Conclusion
Micro-CT is useful for providing non-traumatic, in vivo, quantitative, precise data, and is ideal for exploring the degradation of implants. Regarding the degradation of the magnesium alloy and PLGA in vivo, the rate was slow in the early stage, and then increased with time. The PLGA degrade faster but was then overtaken by AZ31 at week 16. During the degradation, the density of the magnesium had almost no change. The biomaterials were not firmly attached to the surrounding tissues due to inadequate holding forces.
Medicine,Materials Science