Diagnosis of NSCLC on sputum by neutral desorption extractive electrospray ionization tandem mass spectrometry

WEI Yi-ping,YAN Fei-yan,JIA Bin,DING Jian-hua,PENG Jin-hua,CHEN Huan-wen,XU Jian-jun
DOI: https://doi.org/10.3760/cma.j.issn.1001-9030.2011.03.034
2011-01-01
Abstract:Objective To study the diagnosis of non-small-cell lung carcinoma (NSCLC) on sputum analysis by neutral desorption extractive electrospray ionization tandem mass spectrometry (ND-EESIMS).Methods The sputum samples from 6 NSCLC patients and 6 patients without NSCLC were prepared for this approach.Without any preconditioning,home-made neutral desorption extractive electrospray ionization tandem mass spectrometry (ND-EESI-MS) was used for sputum samples analysis.The ND-EESI ion source mode was set to positive ion mode,mass rang 50-1000 Da,Ion transmission tube temperature was set to 150 ℃,atomization voltage was set to 3.5 kV,lens voltage was set to 60.0 V,sample injection tube N2 gas pressure was 3.2 Mpa.The data processing method of principal component analysis (PCA) was performed for the metabolic fingerprinting spectrum.Results Specific sputum sample fingerprint spectrum differences can be observed between the patients with and without NSCLC,such as,78,87,99,118,362,367,539 m/z,the most obviously difference of relative abundance were 362 m/z and 367 m/z.Pattern recognitions at the fingerprint spectrum level for clustering analysis of sputum samples are achieved successfully.Conclusion The patients sputum samples with and without NSCLC can be distinguish effectively by ND-EESI-MS coupling with PCA analysis,the ND-EESI-MS can be a novel methods for rapid and noninvasive diagnosis of NSCLC.