Comparative Study of PCR Kits with Cell Culture and LCR in the Detection of Chlamydia Trachomatis Infections

王宝玺,倪安平,郑和义,朱学骏,秦俭,叶顺章,王千秋,乐嘉豫,汤全贵,刘全忠
2004-01-01
Abstract:Objective: To determine the diagnostic performances of six Chinese PCR kits for detection of Chlamydia trachomatis in patients with sexually transmitted diseases using cell culture and LCR as references. Methods: Endocervical or urethral swab specimens were collected from 673 patients attending STD clinics in Beijing, Shanghai, Nanjing and Tianjin. C.trachomatis culture and PCR were performed on specimens from all patients while LCR detection was performed only on specimens with discordant culture and PCR results. Results: Of the 616 patients, 6.3% (39) were culture-positive while 23.5% to 28.7% were positive by PCR testing. Compared to cell culture, the sensitivity of all six PCR methods was 90% or higher. In 200 cases with discrepant reports, LCR and PCR showed excellent consistency (YI index: 0.523-0.881), the sensitivity and specificity of PCR methods were 83.9%-98.6% and 66.7%-94.7% respectively,while PCR2 showed the highest YI index (0.881). With the reference standard defined as culture positive or LCR positive plus at least one PCR positive for discrepant results, we found that the specificity and sensitivity of all six Chinese PCR kits were higher than 95% and 85%, respectively. Conclusions: Domestically-produced PCR kits for Chlamydia trachomatis detection are highly sensitive and specific, however, quality control remains important in their clinical application.
What problem does this paper attempt to address?