Abstract:OBJECTIVE:To establish a HPLC method for the determine of aceclofenac extended tablets.METHODS:A Zorbax SB-C18 column was used.The mobile phase was acetonitrile-tetrahydrofuran-glacial acetic acid(25∶25∶50,adjusted to pH 3.5 with 1.0 mol/L NaOH) and the flow rate was 1.0ml/min,intertal standard was para-hydroxybiphenyl.The UV detection wavelength was 275nm.RESULTS:The calibration curve was linear in the range from 10.2μg/ml to 50.1μg/ml(r=0.9993) and the average recovey was 100.3%(RSD=0.45%).CONCLUSION:The method is selective,simple and accurate.
What problem does this paper attempt to address?