Removal of Iron Dextran During Hemodialysis with Polysulfone F6 and Homophone GFSplus12

CHEN Yu-qing,WANG Mei
DOI: https://doi.org/10.3969/j.issn.1007-7669.2005.04.013
2005-01-01
Abstract:AIM: To measure the removal of iron dextran in patients undergone hemodialysis. METHODS: Six patients undergone hemodialysis receiving iron dextran 100 mg intravenously for 30 min during h 2 of dialysis. All patients received F6 (polysulphone membrane) and GFSplus12 ( hemophone membrane) dialyzers, separately. A baseline sample of fresh dialysate was collected before the initiation of dialysis. A control sample of spent dialysate was collected during the first hour of dialysis before the infusion of iron dextran. And then, the spent dialysate was again collected from infusing iron dextran until the end of dialysis. The samples were dealed with lanthanum nitrate solution, and then measured by atomic absorptive spectrum method. RESULTS: The iron concentrations at baseline and in control samples were 0.82 and (4.3± s 2.6)μg·L -1. After intravenous infusion of iron dextran, the iron concentration in the spent dialysate increased significantly. The results showed F6 dialyzer ((12±6)μg·L -1, P0.05) or GFSplus12 dialyzer ((15±7)μg·L -1, P0.01) , with minimal amount of iron eliminated through the dialyzers. i.e. less than 3 % of infusion dose outcoming with no obvious difference in the removal of iron for both F6 and GFSplus12 dialyzers. CONCLUSION: The removal of iron dextran by hemodialysis with F6 and GFSplus12 dialysers is minimal demonstrating the efficiency and convenience of the two dialyzers.
What problem does this paper attempt to address?