The Application of CBCT in the Clinical Evaluation of Bone Quality in the Dental Implant Site

Min Peng,Wei Fei
DOI: https://doi.org/10.3969/j.issn.1674-2257.2016.06.007
2016-01-01
Abstract:Objective To investigate the correlation between the gray values of the dental Cone Beam Computed Tomography (CBCT)and the Hounsfield units of multi slice spiral CT (MSCT)and explore the CBCT gray values of the dental implant site and the subjective evaluation of clinical bone quality.Methods The radioactive samples of different concentrations of both dipotassium hydrogen phosphate (K2 HPO4 )and calcium hydroxyapatite (HA)were imaged by both CBCT and MSCT respectively to evaluate the correlation between the gray values of CBCT and HU of MSCT.1 52 patients who were to undergo posterior dental implant were scanned by CBCT before the operation,and the gray values of the bone in the dental implant site were measured and compared with the dentist′s subjective evaluation of the bone quality.Results The gray values of CBCT were highly correlated with HU of MSCT.There was a decreasing trend in the gray values of CBCT in the dental implant sites in accordance with the bone types,and the mean gray values of the four bone types were 312 (Type 1),214 (Type 2),76 (Type 3),and -454 (type 4)respectively.Conclusion The gray values of CBCT can be applied to evaluate the bone density and provide valuable references for predicting the bone quality in the potential dental implant site.