Comparison of virtual touch tissue quantification for measurements of liver stiffness bteween convex and linear ultrasound transducer

刘卫勇,隋秀芳,康冰飞,刘力,单立奎,叶磊
DOI: https://doi.org/10.19405/j.cnki.issn1000-1492.2013.12.031
2013-01-01
Abstract:Shear wave velocity (SWV) was performed in 31 patients with chronic hepatitis B (CHBs) and 45 healthy volunteers (control group). The occurrence rate of absence data with 4C1 was lower significantly than with 9L4 transducer both in CHBs and control group. The occurrence rate of absence data in the CHBs was not different from that in healthy controls with 4C1 transducer, which also could be seen when using 9L4 transducer. The SWV were significantly higher with 9L4 than 4C1 transducer both in control group and in CHBs. For SWV measurements performed by 4C1 and 9L4 transducer, the intraclass correlation coefficient (ICC) in the CHBs and control group were classified as poor and fair agreement. Bland-Altman analysis compared the differences against mean of SWV measured by 4C1 and 9L4 transducer for the CHBs and control group were at -0.85 and -0.64, respectively. VTQ in its present stage of development has low agreement between 4C1 and 9L4 transducer in the assessment of liver stiffness.