Coronary Stent Implantation in Canine Model and Technique Approach

何奔,郭伟峰,赵军礼,周国伟,孙宝贵
DOI: https://doi.org/10.3969/j.issn.1674-8115.2004.z1.011
2004-01-01
Abstract:Objective To established a stable coronary artery stent implantation model in canine and discuss the techniques. Methods Coronary artery stent implantation techniques in canine were accomplished via right carotid artery access by DSA. Results 25 canines except 1 died of inappropriate anaesthesia were accomplished stent implantation and 19 of them survived long enough for further study. In 3 of 24 canines who were successfully implanted stent, the wire in aortic arc pointed to descending aorta and need a left Judkin's catheter to introduce wire to ascending aorta. Of 24 canines, right coronary Judkins guiding catheter(GC) were well engaged in 17 canines, 7 cases needed heat reshapable GC or in half engagement state manipulate wire into coronary artery skillfully then introduce GC well cannulated. GC pointed to LAD in 9 cases and pointed to LCX in 15 cases. The mean fluoroscope time were 18 min in first 12 cases and 11 min in later 12 cases. The mean contrast consumption were 90 ml in first 12 cases and 78 mL in later 12 cases. Stent implanted in LAD in 11 cases and LCX in 13 cases, all of them were direct stenting without predilatation. 9 of them were premounted stents and 15 of them were mounted to balloon by operator and none of stent dislodgement occurred. Conclusion Stent implantation in canine coronary artery via carotid approach can be made with high successful rate and the key point to success is well good GC engagement.