Quantitative Measurement and Stability Study of Betaine and Free Choline in Human Plasma by Liquid Chromatography Tandem Mass Spectrometry

ZHANG Zhen-zhen,WANG Jia-jia,TAN Bing-yan,WANG Xian,ZHENG Lin,CHANG Ying,ZHU Hui-lian
DOI: https://doi.org/10.13471/j.cnki.j.sun.yat-sen.univ(med.sci).2011.0055
2011-01-01
Abstract:[Objective]To establish a method for the quantitative measurement of betaine and free choline in plasma simultaneously using liquid chromatography tandem mass spectrometry(HPLC-MS/MS) and to study the stability of the two analytes in plasma.[Methods] Separation was achieved on a normal-phase silica column(2.1 mm × 100 mm,5 μm) with acetonitrile/15 mmol/L formic aicd(6:4v/v) under isocratic elution.A mass spectrometry was applied to measure the objective analytes.Acetonitrile,contained d9-choline and d9-betaine,as internal standards,was added into plasma to precipitate protein.[Results] For both betaine and choline,the lower limit of quantitation(LLOQ) was 0.62 μmol/L and the response was linerar over the range investigated(0.62 μmol/L-400.00 μmol/L,r2 0.999).The within-day relative standard devition(RSD) were below 8% and between-day below 15%.The recoveries were over 90%.For the stability tests,the concentration of the two analytes increased with the storage temperature and time.Choline was stable throughout three freeze-thawn cycles while concentration of betaine had a significantly rise after the first cycle.[Conclusion] This is a technically simple and accurate method that can be carried out with HPLC-MS/MS.Attention should be paid to the critical effects of storage temperature and time and freeze-thawn cycle on the measurement.
What problem does this paper attempt to address?