Improvement of Determination of Camphor and Eucalyptol in Shidishui Tincture by Gas Chromatography

Yuanyuan LIU,Shihai YU,Jia HU,Tao HE
DOI: https://doi.org/10.13847/j.cnki.lnmu.2014.05.002
2014-01-01
Abstract:Objective To establish the student experimental method of measuring the content of camphor and eucalyptol in Shidishui tincture by gas chromatography. Methods SE-54 capillary column (30 m×0. 25 mm×0. 25 μm) was used in the experi-ment. The injector temperature was 230 ℃ , and FID detector temperature was 240 ℃ . The Temperature was programmed with the ini-tial 70 ℃ and the final 160 ℃ , increasing at the speed of 8 ℃ ·min-1 . Results The average sample recovery rates of camphor and eucalyptol in the experiment were 98. 3% and 97. 6% respectively, and RSD were 1. 9% and 2. 1% respectively. Conclusion The method can be applied to student experiment teaching for its simplicity and high accuracy.
What problem does this paper attempt to address?