Experimental Study with Domestic Bone Cement in the Percutaneous Vertebroplasty

倪才方,刘希胜,陈珑,杨惠林,唐天驷,丁乙
DOI: https://doi.org/10.3969/j.issn.1673-0399.2004.05.004
2004-01-01
Abstract:Objective To study the feasibility of injecting domestic bone cement in the process of the percutaneous vertebroplasty. Methods (1)Various types cement were disposed with domestic PM-MA. The concretionary phases of cement were observed according to the stages and holding time. Then the most ideal ratio of the mixed cement was selected and ten cement columns were made with this ratio, which was taken as the trial group. The other ten referring to was taken as the contrast one. The biological mechanics was measured with a load and the data of the results were compared. (2) Twenty thoracic and lumbar adjacent bodies were removed intact from five human corpses. These bodies were divided into two groups, in one group PMMA were injected, the other was severed as the contrast one. Then in these twenty vertebral bodies the biological mechanics was measured and the treatment effect was evaluated. (3)In 12 healthy dogs PVP in lumber was tried so as to observe the operational difficulty during the process of injected this bone cement and CT was used to evaluate the result of PMMA, diffusion and the complications caused by it. Results The most ideal ratio was 4:2.6 :1(g, ml, ml) between powder, monome and contrast. After injecting this kind of cement, the loading strenght of these vertebral bodies was increased remarkably (P0. 01). Conclusion Injecting domestic bone cement provides the theoretical foundations for the clinical application of PVP.
What problem does this paper attempt to address?