Effect of new ultrasonic catheter on disruption of thrombi in vitro

Jiayi TONG,Qimin DAI,AG Mrochek,VI Stelmashok,VT Minchenya
DOI: https://doi.org/10.19460/j.cnki.0253-3685.2005.10.007
2005-01-01
Abstract:Objective To assess the efficacy of new ultrasonic catheter for thrombus dissolution in vitro.Methods 54 samples of human whole blood thrombi were studied in vitro.Thrombi were disrupted with three kinds of ultrasonic catheters,which designed with different tips,ball-tip,ball-tip with a truncated top or ball-tip with indentation.Results All three types of catheters effectively disrupted the thrombi.Ultrasonic catheter with ball-tip and ball-tip with indentation had better effect on disrupting thrombi than ultrasonic ball-tip with a truncated top.Ninety percent residual particles were smaller than erythrocytes in diameter.Conclusion The high-intensity,low-frequency catheter-delivered ultrasound can be effective and relatively safe in ablating thrombi.In vitro experimental findings suggest that modification of ultrasonic tip has a marked influence on the ability of whole blood thrombi disruption.
What problem does this paper attempt to address?