Determination of contents of fingolimod in fingolimod capsules by HPLC

Wei Wang,Haibin Lu
DOI: https://doi.org/10.3969/j.issn.1004-0188.2014.10.005
2014-01-01
Abstract:Objective To establish a method for determining the contents of fingolimod in fingolimod capsules by High Performance Liquid Chromatography( HPLC). Methods HPLC was used to determine the contents of fingolimod in fingolimod capsules. The selected column was Agilent TC-C18,and the mobile phase was sodium perchlorate solution(0. 1 mol/ L aqueous solution of sodium perchlorate,and the pH value of perchloric acid had been adjusted to 2. 8):methanol :acetonitrile(93:7:100);the detection wavelength was 215 nm with a flow rate of 1. 5 ml/ min;the sample size is 10 μl. Results The linear range of concentration of fingolimod was 0. 0499 ~ 0. 1499 mg/ ml(r = 0. 9998);the average recovery rate of samples was 100. 01%(RSD = 0. 46% ,n = 3). Conclusion This method is simple with high reproducibility and high accuracy,and can be used to determine the contents of fingolimod in fingolimod capsules for quality control.