Determination nitrogen in the Kjeldahl digests of plant samples by continuous flow analyzer in comparison with auto-mated distillation-titration instrument

温云杰,李桂花,黄金莉,刘云霞,高翔,汪洪
DOI: https://doi.org/10.11838/sfsc.20150624
2015-01-01
Abstract:Kjeldahl distillation-titration has been used as a reference method for N determination in plants, but it is time consu-ming. This article aimed at establishing a time saving method for determining of plant N concentrations by continuous flow ana-lyzer ( CFA) . Twenty four of wheat plant samples were selected and digested with H2 SO4-H2 O2 . N contents in the digests were determined by CFA and an automated Kjeldahl distillation-titration ( AKDT) instrument, respectively. The test showed that there was no any significant difference of the plant N contents measured by the CFA and the AKDT method. A linear rela-tionship best described the measured N values generated by both methods: CFA-N=0. 892 AKDT-N +0. 753. The Pearson’s correlation coefficient was 0. 942 1 with a significance level (n=24, P<0. 01). The CFA method for plant N measurement had a high precision with relative standard deviation less than 5%. The standard recovery rate of feed samples with addition of ( NH4 ) 2 SO4 was 96. 6% ~102. 3%. It suggested that the CFA based on colorimetric detection of the pure ammonium chloride distillate ( Berthelot reaction) could be used to determinate N in the Kjeldahl digests of plant samples. The CFA method was labor saving and could handle large numbers of samples. The human error in mixing and other operations was reduced to a great extent.