A calibration transfer method for NIR model based on extended spectrum

Jiajun WANG,Wei ZHE,Yan LIU,Wensheng CAI,Xueguang SHAO
DOI: https://doi.org/10.3969/j.issn.1004-5708.2014.06.001
2014-01-01
Abstract:A method for calibration transfer was proposed based on extended spectra of a small standard sample set between master and slave machine. By using the variance spectra, a large number of spectra measured on master machine were transferred to the spectra of slave machine, which was named extended spectra. Model of slave machine was built with the extended spectra as a calibration set. Near infrared spectra on three different instruments were investigated to validate the model. Results showed that an accurate model could be obtained with the extended spectra calculated from a small set of standard samples. The method provided a useful tool for model updating when changing the mechines in practical uses.
What problem does this paper attempt to address?