Evaluation and Application of a Quantitative Isothermal Amplification Fluorescence Kit for Rapid Detection of Salmonella

Hao-xuan TANG,Xue-yang SHAO,Pei-en NI,Ming-sheng TIAN,Zhi ZHU,Rong-yong YIN,Shao-shui ZHONG,Qing-li DONG,Geng-sheng HE,Yuan-qi FENG,Da-peng WANG
DOI: https://doi.org/10.3969/J.ISSN.1671-9964.2017.06.013
2017-01-01
Abstract:Rapid detection of foodborne bacteria is a key for food safety.Salmonella is one of the important foodborne pathogenic bacteria.In this study,the artificially contaminated samples with Salmonella were used to evaluate the sensitivity and accuracy of a retail kit based on isothermal amplification.Three kinds of samples were detected,including bacteria culture,food and clinical samples.The results demonstrated that the detection limit of the kit was 103 CFU/ml and the accuracy rate was over 96.7%.In addition,the false-positive rate in all negative samples was 0.0 %,3.3 % and 3.3 % respectively.The positive rate in all positive samples was 97.9%,100.0% and 65.7% (including low detection limit samples)respectively.In conclusion,the advantage of the kit was less time consuming,simple protocol and wide adaptability.