Design of Single-Cell Sensor Based on Light-Addressable Potentiometric Sensor
QIN Li-feng,XU Gai-xia,LI Rong,YE Xue-song,WANG Ping
DOI: https://doi.org/10.3785/j.issn.1008-973x.2005.09.027
2005-01-01
Abstract:To measure neural electrophysiological characteristics and to evaluate drugs, a novel cell-based biosensor based on light-addressable potentiometric sensor (LAPS) was proposed. The sensor sensitivity was detected by adding a sine wave signal with amplitude of 2 mV and frequency of 1 kHz as simulated neural action potential (AP). Neurons were immobilized on the sensor surface by depositing bio-layers and changing surface roughness to improve the sensitivity. Acetylcholine (Ach) with concentration of 1 mg/L, 10 mg/L was chosen as stimulating drug of cortical neuron, and the AP of the neuron was detected under different drug concentrations. The results show that surface modification can improve the sensor sensitivity, and the proposed sensor can successfully detect the simulated AP signal. The single-cell sensor is able to distinguish the response of neurons under different drug concentrations.
What problem does this paper attempt to address?