A quantitative determination system for fluorescent lateral flow immunochromatographic strip based on STM32

Haiyan JIANG,Hongzhou HUANG,Yueming GAO,Jianguo CHEN
DOI: https://doi.org/10.13764/j.cnki.ncdg.2017.01.017
2017-01-01
Abstract:This paper introduced the design of quantitative detector for fluorescence lateral flow immunochromatographic strip with STM32 as a micro controller,which can realize the quantitative detection on C-reactive protein.The system was composed of a set of modules,including optical module,photoelectric sensor,and fluorescent signal acquisition,conditioning circuit,LCD display and mechanical scanning devices.The experimental test showed that the lower limit of detection (LOD) was 1.95 μg·mL-1,and the linear range was 1.95-256 μg·mL-1 for C-reactive protein (CRP) in the system.It has robust linearity and stability and can be utilized widely in different fields,such as biomedical detection,clinical application and environment determination.