The Progress of Joule Balance

Zhong-hua ZHANG,Chen LI,Qing HE,Zheng-kun LI,Bing HAN,Yun-feng LU,Jiang LAN,Shi-song LI,Ya-qiong FU,Gang WANG
DOI: https://doi.org/10.3969/j.issn.1000-1158.2014.04.01
2014-01-01
Abstract:The National Institute of Metrology(NIM)has been studying the Joule Balance method(which is used for the same purpose as the Watt Balance method but operated differently),for several years to establish a kind of quantum mass standard by means of measuring the Planck constant. Since 2010,some progress has been achieved. A new standard square wave compensation method has been developed to measure the DC mutual inductance with an uncertainty of several parts in 10 - 7(k = 1). . An eddy current sensor for positioning has been designed and implemented to obtain a more accurate position of the movable coil. . The alignment technique for the coil system has also been placed under consideration. Specifically,a newly developed oil damping device reduces the small swing of the coil set(which occurs because of a combination of the weak vibration of the balance base and the hot ascending warm air from the current heating of coils). As a result,the measurement uncertainty of the coil position and mutual inductance is improved. . The recent measurement result of the Planck constant h is 6. 626104(59)× 10 - 34 J·s with an uncertainty of 8. 9 × 10 - 6( k = 1). Further improvements and difficulties are also discussed.
What problem does this paper attempt to address?