Analytical method and performance research for determination of hydrogen and oxygen stable isotopes ratio in water sample by using new type of laser spectrum analyzer

Biao TIAN,Weijun SUN,Minghu DING,Tong ZHANG,Cunde XIAO,Dongqi ZHANG
DOI: https://doi.org/10.11857/j.issn.1674-5124.2018.02.025
2018-01-01
Abstract:In order to understand performance characteristics and test the accuracy of laser isotope spectrometer and reduce its memory error, memory error test have been carried out for rapid analysis method and high-precision analysis method of Picarro L1102i laser isotope spectrometer which is based on the wavelength-scanned cavity ring down spectroscopy (WS-CRDS)technology, and the accuracy comparison was carried out with the test results of MAT 253 isotope ratio mass spectrometer so as to verify the reliability of the method. Results show that accuracy of data of high-precision analysis method is superior to rapid analysis method but there is no essential difference between them. The average deviation of δ18O test value and MAT 253 test value obtained from these two methods is below 0.000 1% and the average deviation of δD test value and MAT 253 test value is respectively 0.001 6% and 0.000 6%, both of them meet precision test requirements. The preparation process of sample based on Picarro L1102i spectrometer is simple and its test cost is lower than that of MAT 253 mass spectrometer, which is greatly potential in routine water sample analysis.