Magnet-coil Force Actuator for Microtensile Test Device

Ding Jianning,Meng Yonggang,Wen Shizhu
DOI: https://doi.org/10.3321/j.issn:0254-3087.2000.05.001
2000-01-01
Abstract:In order to test and evaluate the mechanical properties of the used films or microstructures accuratly, the optimum configuration and geometry for a magnet-solenoid force actuator have been presented, enabling its application to the design of microtensile test device. In order to precision controlled, the linear relationship between the magnetic field gradient and the coil excitation current has been achieved. At last, the microtensile test device is described. One end of the sample is attached to a movable rod suspended by leaf springs and the other end is held fixed. Force is applied to the rod by a coaxial electromagnet acting on permanent magnets attached to the rod. Displacements are measured with optical fibers. For the coil of a1=0.008m,α=2,β=1,χ=0.77 and n=12800, the force on the magnets is linear in current with a slope of ΔF/ΔI=0.503±0.01mN/mA.
What problem does this paper attempt to address?