The fluorescence signal intensity correction of biochip base on galvanometer scanning

Piyin YANG,Zhenhua GAN,Yueming GAO
DOI: https://doi.org/10.3969/j.issn.1005-5630.2017.02.013
2017-01-01
Abstract:Since the scanning speed of the confocal device is slow and structure of the traditional biological chip detection device is complex,we design a new device with laser light source,two-dimensional high speed vibrating mirrors and a cooling CCD camera.Further,a corresponding mathematical test model was developed.The proposed device uses STM32 as the control chip and collects fluorescent signal by scanning the micro-slide in an arched direction.In the experiment,cyanine (Cy5) dye dots in the same micro-slide were scanned in the five different positions.Furthermore,pictures were collected by the cooling CCD camera.The results show that gray images obtained in the five different positions were corrected by detection mathematic model.In addition,gray correction error was within 2%,which satisfies the requirement of light intensity consistency.