An Experimental Study on Determination of the Water Content in Bio-Oil by Karl-Fischer Titration

WANG Hua,LIU Rong-hou,ZHANG Chun-mei,LUAN Jing-de
DOI: https://doi.org/10.3969/j.issn.1671-5292.2005.03.006
IF: 8.7
2005-01-01
Renewable Energy
Abstract:Karl-Fisher titration was used to Determinate the water content of bio-oil in a 758KFD Karl-Fisher titration. Effects of different solutions and bio-oil sample amount on water content in bio-oil were conducted. In addition, water adding in bio-oil sample experiments were conducted to test the accuracy and stability of the method. From the results, it can be seen that the water content in bio-oil has less changes,which were 36.86wt% and 36.16wt% respectively, when methanol or the solution of the mixture of chloroform and methanol(3∶1) were used. Methanol can be selected as test solution. The sample amount has no more influence on water content in bio-oil. When sample amount of bio-oil was 0.06 g, the water content was stable, averaging 42.63wt%. The measuring water content in bio-oil is a little higher than that of calculation result drived from water adding in bio-oil sample experiments.
What problem does this paper attempt to address?