Use of radiochromic film for diagnosis of accelerator beam position

Yaqian CAI,Qingxi YANG,Kaizhong DING,Yuntao SONG,Yucheng WU,Lingli MAO
DOI: https://doi.org/10.11889/j.0253‐3219.2019.hjs.42.010202
2019-01-01
Nuclear Techniques
Abstract:According to the beam characteristics and debugging requirements of SC200 superconducting proton cyclotron, it is necessary to measure the beam position accurately in the process of proton acceleration and extraction. [Purpose] This study aims to propose a beam position diagnostic method by means of radiochromic film. [Methods] First of all optimal type of the film was selected by dose calculation at different measuring positions. The feasibility of four types of films was verified by Monte Carlo simulation calculation and experimental measure on ion source test platform. Then the optimal irradiation time for different types of films at different measurement positions was provided through the experiment and dose analysis. Finally, a processing scheme for visualizing the discoloration area of the film was proposed. [Result & Conclusion] The four radiochromic films can be colored after proton irradiation, and the color of the films can be maintained stable after long time, hence can be used for beam position diagnostics.