Trabecular Bone Spacing Estimation Based on Hilbert Transform and Fundamental Fre-quency Estimation Method

Dean Ta,Ying Li,Chengcheng Liu
DOI: https://doi.org/10.16337/j.1004-9037.2015.02.009
2015-01-01
Abstract:Ultrasonic backscatter signal is quite sensitive to the microstructure of cancellous bone.Tra-becular bone spacing (TbSp)is an important parameter for characterizing bone microstructures.In order to acquire TbSp accurately from the ultrasonic backscatter signal of the cancellous bone,a TbSp estima-tion method is proposed using a combination of the Hilbert transform and fundamental frequency estima-tion (HFE)method.The HFE results from the cancellous bone in vitro are compared with the TbSp ob-tained fromμ-CT.The HFE results are accurate (estimation error<3%)and stable (standard variation<4%)at higher frequencies (5 MHz and 10 MHz),and more accurate when standard TbSp is large,and have a high correlation (r 2 =0.75—0.99,p <0.01 ,n=1 6)with the standard TbSp at different frequen-cies.It shows that the HFE method is accurate and stable for TbSp estimation.The HFE method is ver-tified to characterize the cancellous bone TbSp.