Two-dimension automatic tracking vision autocollimator system
Shi-zhe CHEN,Zhao-bang PU,Guo-dong LIU,Wen-qian WANG,Bing-guo LIU
DOI: https://doi.org/10.3969/j.issn.1003-501X.2005.05.015
2005-01-01
Abstract:In order to increase the measurement precision of an autocollimator and decrease the fatigue of eye during aiming target and visual error, a two-dimension automatic tracking vision autocollimator based on machine vision is presented. The system is composed of an optical autocollimator, a CCD and image collector board etc. The system carries out automatic and precision measurement of small angles by using CCD rather than eye to aim cross reticule image on the autocollimator, and by means of auto-search algorithm with image moment invariants, Hough transformation, sub-pixel subdivision, etc. Experiments demonstrate that this system has large measurement range up to ±15' with high resolution of 0.1″ The error in contrast to the MJS double-frequency laser goniometric system is less than 0.6″. This system has higher precision than MJS with merits of convenient operation and easy popularization.
What problem does this paper attempt to address?