An Electrochemical Immunosensor for Rapid Detection of Luteinizing Hormone

Zhuang Kong,Yang Wang,Juntao Liu,Jinping Luo,Hongyan Jin,Xinxia Cai
DOI: https://doi.org/10.13250/j.cnki.wndz.2017.08.004
2017-01-01
Abstract:An electrochemical immunosensor for detection of luteinizing hormone (LH) antigen was developed to realize the high sensitive and rapid detection.With the paper-based immunosensor as the base material,the screen-printed working electrode (SPWE) was modified by the amino functional graphene (NH2-G) /thionine/gold nanoparticles.The detection principle of the immunosensor is that the electrochemical material thionine produces the electrochemical current through the redox reaction of the electrode surface,and the specific binding of the antigen antibody will affect the release of the electrochemical current to decrease the current.The detect current is inversely proportional to the concentration of the antigen.The experimental results show that the sensor has good test results for luteinizing hormone (LH) in the range of 1-100 mIU · mL-1,the lowest limits is 1 mIU · mL-1,the linear coefficients is 0.991,and the detection time can be controlled within 10 min.Therefore,the immunosensor can be used for rapid and high sensitivity detection of luteinizing hormone.Thus,it has an important significance to dynamically monitor the sex hormone levels,then realize personalized diagnosis and treatment of infertility disease and guide the preparation for pregnancy.
What problem does this paper attempt to address?