Improvement and Experimental Research on Magnetic Localization Algorithm of Capsule Endoscopy

Xiu-quan LIU,Yan-hong LI,Chang LIU,Ping HUANG
DOI: https://doi.org/10.3969/j.issn.1000-8829.2017.08.035
2017-01-01
Abstract:In order to solve the problems of complex magnetic field calculation model and poor real-time performance,a backstepping magnetic localization algorithm is put forward.The mathematical model of the magnetic positioning system is established based on the magnetic dipole model.A rapid positioning algorithm is adopted,which constructs the magnetic positioning objective function.According to the output voltage of the magnetic resistance sensors,an initial value of the position and direction parameters of the capsule endoscopy permanent magnet is calculated by the linear algorithm,the optimal solutions of the capsule endoscopy permanent magnet are derived by the nonlinear iterative algorithm,the complexity of the localization algorithm is simplified,and the real-time positioning is improved.The endoscope magnetic positioning system is established,the results show that the design of the capsule positioning system is feasible and effective.
What problem does this paper attempt to address?