Proficiency testing on determination of 6 kinds of liver function indices

Yu-Fa MIAO,Yue GU,Chao WANG,Lin ZHANG,Dong-Sheng PAN,Xue WANG,Jian-Jun LV,Zhi-Ying HUANG,He-Zhan ZHANG
DOI: https://doi.org/10.3969/j.issn.2095-0381.2018.07.023
2018-01-01
Abstract:Objective To evaluate the ability for detection 6 kinds of liver function indices by Chinese institutions of food and drug control, including alanine aminotransferase (ALT), aspartate animotransferase (AST), alkaline phosphatase (ALP), total bilirubin (TBIL), total protein (TP) and albumin (ALB). Methods The uniformity and stability of proficiency testing samples were analyzed by one-way ANOVA and t test, respectively. Normality test was performed for data distribution, and the proficiency testing results were evaluated according to evaluation criterias. Results The satisfaction rate of ALT was 97. 5%, the satisfaction rate of AST was 97. 6%, the satisfaction rate of ALP was 100%, the satisfaction rate of TBIL was 74. 2%, the satisfaction rate of TP was 90% and the satisfaction rate of ALB was 97. 5%, respectively. Conclusion Through this study, the detection ability defects of liver function indices were discovered in several laboratories, thus it promoted the quality improvement of the laboratory.