Determination of riboflavin in beverage by combination of standard addition method and a mobile phone controlled fluorescence detection system

Zhong-Rong WANG,Qing-Jie FAN,Yun-Fang JIA
DOI: https://doi.org/10.3969/j.issn.2095-0381.2017.11.039
2017-01-01
Abstract:Objective To establish a method for determination of riboflavin in dietary beverage by combination of the proposed mobile phone controlled fluorescence detection system and standard addition method. Methods The fluorescence signals of riboflavin in 3 kinds of beverages (beer, orange juice and apple vinegar) were detected by the proposed mobile phone controlled fluorescence detection system. The concentrations of them were determined by standard addition method. Results The riboflavin working curve of this system had a good linear relationship, and the linear equation was Y=0.1857X+0.2133 with R2of 0.9649. The concentrations of riboflavin in beer, orange juice and apple vinegar were 2.06, 4.58, and 0.28 μg/mL, respectively, which were in agreement with the results determined by Vitamin Detector LK3000V. Conclusion The combination of standard addition method and the proposed mobile phone controlled fluorescence detection system can realize the detection of riboflavin in beverage, which provides an effective and easy-to-operate riboflavin detection equipment for food analysis.
What problem does this paper attempt to address?