Determination of Caffeine in Tea Drinks by HPLC

Xian-qing HUANG
DOI: https://doi.org/10.3969/j.issn.1007-7871.2015.01.012
2015-01-01
Abstract:HPLC is used to explore the factors such as column temperature, wavelength, mobile phase and flow rate’s of caffeine in tea. The best conditions is wavelength of 273nm, mobile phase acetonitrile: water 20∶80, column temperature 20℃, and the flow rate of 1mL/min. In concentration range of 0~200μg/mL, linear relation was good. Standard addition recovery rate was 99.69%. Established methods can accurately determine caffeine in tea drinks.
What problem does this paper attempt to address?