Study on the Analytical Method of Some Heavy Metal Ions with Diethuldithiocarbamate Chelates by the Reversed-Phase High Performance Liquid Chromatography

张莉,张国英,崔兆杰
DOI: https://doi.org/10.3969/j.issn.1002-6002.2005.01.011
2005-01-01
Abstract:The newly method of Reversed_phase High Performance Liquid Chromatography has been set up and applied to the qualitative separation and quantitative determination of some heavy metal ions including Cu(Ⅱ),Hg(Ⅱ),Pb(Ⅱ),Zn(Ⅱ),Cd(Ⅱ),Cr(VI) by using the diethyldithiocarbamate as the chelates and chloroform as the extractive solvent in the paper.The researching results show these heavy metal ions can be separated effectively by using acetonitrile∶methanol∶water=55∶15∶30 as mobile phase and C18 reversed_phase column(5μm,1 50×4.6mm i.d.),under the flow_rate of 1.0mL/min and detection wavelength of 265nm,the detection limit is from 0.1 ng to 1.6ng,the linearity range and relation can meet the requirement ofquantitative analysis for the trace heavy metal ions,the method has good reproducibility and sensitivity.
What problem does this paper attempt to address?