Determination of Benzo(a) -pyrene in Soil by Ultrasound Extraction and High Performance Liquid Chromatography

Ying Zhu,Pei Liu,Dilan Qin,Bingbing Song,LuoYueping,Junping Bi,Rilong Zhu
DOI: https://doi.org/10.3969/j.issn.1673-1212.2016.04.028
2016-01-01
Abstract:A novel method for detection and determination ofbenzo( a) -pyrenein soil was developed, using ultrasound ex-traction-HPLC. Soil samples were extracted with hexane/dichloromethane(1+1)ultrasonic extraction, extract by silica gel col-umn or Florisil silica column purification for volume and concentration, acetonitrile water as mobile phase, linear gradient elution, a flow rate of 1. 0 ml/min, fluorescence detector excitation wavelength was at 280 nm and emission wavelength of 420 nm, exter-nal standard method and peak area quantification. . Result showed that the limit of detection for benzo( a) -pyrene was0. 15μg/kg. The method can accurately determine benzo( a) -pyrene, the relative standard deviation was in the range of 3. 9% ~8. 4%. The real sample results showed that the recovery ratio of standard addition was in the range of 72. 5% ~85. 8%. Therefore the method can well satisfy the need of practical determination in soil.
What problem does this paper attempt to address?