Investigation Of Pet Imaging Experiment And Monte Carlo Simulation For Cancer Therapy Using Heavy-Ion Beams At Hirfl

Zq Chen,Yl Ye,C Li,Yf Yang,Wl Zhan,Gq Xiao,Hs Xu,Zy Guo,Wf Li,Dx Jiang,Qj Wang,T Zheng,Gl Zhang,Zh Li,Xq Li,Qy Hu,Dy Pang,W Jia
DOI: https://doi.org/10.3321/j.issn:0254-3052.2004.07.022
2004-01-01
Abstract:For cancer therapy using heavy-ion beams at HIRFL, a simple Positron Emission Tomography(PET) imaging system was made using two position sensitive BGO scintillator detectors. The object images and gamma ray energy spectra were measured for the PET. The energy resolution for the 511 keV full energy peak is 18.6% and its countiong fraction is 52.4%. Monte Carlo calculations were performed for the PET imaging system by using GEANT3 code. The object images and I ray energy spectra are in good agreement with the experimental data. In order to develop a PET imaging system with high efficiency, good resolution and low cost, the PET imaging system was optimized.