Accurate Determination of Cross-Sections for 238U (n, 2n) 237U Induced by Neutrons Around 14 MeV

Xianggao Wang,Shan Jiang,Ming He,Kejun Dong,Caijing Xiao
DOI: https://doi.org/10.1016/j.nima.2010.04.034
IF: 1.335
2010-01-01
Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment
Abstract:The cross-sections of 238U (n, 2n) 237U induced by neutrons with energies around 14MeV were determined by measuring the emission rate of 208.00keV γ-rays from 237U decay with an accurately calibrated HPGe γ-ray spectrometer at CIAE. In order to correct for the self-attenuation of 208.0keV γ-ray in U sample, a 177Lu-spiking method was introduced. As a result, a total relative standard uncertainty of 2.5% was achieved for the measured cross-section values.