TB PCR in BAL and EBUS-TBNA samples for the diagnosis of pulmonary and mediastinal lymph node TB: retrospective TRiBE study

Mirae Park,Kartik Kumar,Meg Coleman,Laura Martin,Georgina Russell,Pauline Scheelbeek,Ajit Lalvani,Giovanni Satta,Onn Min Kon
DOI: https://doi.org/10.1136/thorax-2023-220647
2024-08-19
Thorax
Abstract:Introduction The role of Xpert Ultra in bronchoalveolar lavage (BAL) and endobronchial ultrasound-guided transbronchial needle aspiration (EBUS-TBNA) samples for pulmonary and mediastinal lymph node tuberculosis (TB) remains unclear. Methods This was a retrospective observational service evaluation at a tertiary TB centre in a low-incidence setting. The diagnostic indices of Xpert Ultra, smear and culture (with cytology for EBUS-TBNA samples) were compared with culture positivity or a composite reference standard of clinical TB diagnosis. Trace readouts, a new category of results for Xpert Ultra indicating low bacillary load, were analysed in two ways as a true positive or true negative result. 282 BAL and 139 EBUS-TBNA samples were included in the analysis. Results BAL: sensitivity with 95% CI against culture-confirmed pulmonary TB from BAL samples for Xpert Ultra (trace as positive) was 0.91 (0.82 to 0.98), Xpert Ultra (trace as negative) was 0.76 (0.69 to 0.83), smear was 0.38 (p=0.0009) and culture was 1.00 (0.91 to 1.00). Specificities for all the tests were ≥0.99 (0.98 to 1.00). The addition of smear to Xpert Ultra did not improve the diagnostic accuracy. EBUS-TBNA: sensitivity against culture-confirmed TB from EBUS-TBNA samples for Xpert Ultra (trace as positive) was 0.71 (0.63 to 0.78), Xpert Ultra (trace as negative) was 0.59 (0.54 to 0.63), smear was 0.12 (p=0.002), culture was 1.00 (0.89 to 1.00), cytology was 0.87 (0.76 to 0.98) and rapid on-site evaluation of cytology (ROSE) was 0.92 (0.78 to 1.00). Specificities were 0.99 (0.97 to 1.00), 0.99 (0.97 to 1.00), 1.00 (0.98 to 1.00), 1.00 (0.98 to 1.00), 0.67 (0.67 to 0.68) and 0.42, respectively. Conclusion Xpert Ultra had a significantly higher sensitivity compared with smear in both BAL and EBUS-TBNA samples. Xpert Ultra had a lower sensitivity compared with culture but comparable specificity with results being available within <24 hours. Trace readings in our low-incidence setting were associated with culture positivity in all BAL samples.
respiratory system
What problem does this paper attempt to address?