Magnetic Resistance Sensory System for the Quantitative Measurement of Morphine

Zhengkang Chu,Miaomiao Fu,Jiuchuan Guo,Wei Wang,Jun Zhou,Xing Ma,Jinhong Guo
DOI: https://doi.org/10.1109/TBCAS.2021.3060181
2021-01-01
IEEE Transactions on Biomedical Circuits and Systems
Abstract:Point-of-care testing (POCT) is characterized by fast detection, simple use, and cost efficiency. As the focus of healthcare shifts to precision medicine, population health, and chronic disease management, the potential impact of POCT has increased significantly in the past decade. Immunochromatographic test strips (ICTSs) are currently the most promising POCT diagnostic format due to the advantages of fast detection, simple operation and cost-effective. The lateral flow immune analyzer (LFIA) system that uses magnetic resistance (MR) sensors as readers and magnetic nanoparticle labeling materials has attracted wide attention due to its high sensitivity and specificity. This paper designs a quantitative lateral flow immunoassay instrument for MR sensors using InSb single crystals as magnetoresistive elements. MR sensors detect samples of ICTSs labeled with superparamagnetic nanoparticles (SMNPs). The collected weak magnetic signal is processed by the designed amplifying circuit and software algorithm. As a result, the MR system can quantitatively detect the concentration of morphine (MOP), and has good repeatability. Compared with other optical detection instruments, the system has good specificity because there is very low background noise in the detection sample. The size of the whole system is 98 mm x 62 mm x 78 mm, and it has the characteristics of compactness and fast detection.
What problem does this paper attempt to address?