Hierarchical Porous Tubular Biochar Based Sensor for Detection of Trace Lead (II)

Xue Chen,Kunchao Lu,Donghai Lin,Yan Li,Shiyu Yin,Zhiyi Zhang,Meihua Tang,Guosong Chen
DOI: https://doi.org/10.1002/elan.202060148
IF: 3.077
2021-01-01
Electroanalysis
Abstract:Hierarchical porous tubular biochar (PTBC) was prepared by selectively removing lignin simply according to reverse the pyrolysis sequence of cellulose. The properties of the PTBC sample were characterized by XRD, SEM, TEM, Raman spectra, cyclic voltammetry and electrochemical impedance spectroscopy. The electrochemical performances of PTBC modified on the screen-printing electrode illustrated excellent detection of lead ions (lead (II)) in water with the linear range (0.5-120 mu g/L) and the detection limit (0.02 mu g/L) by in-situ bismuth film square wave anode stripping voltammetry.
What problem does this paper attempt to address?