An Amperometric Hydrogen Peroxide Sensor Based on Reduced Graphene Oxide/Carbon Nanotubes/Pt NPs Modified Glassy Carbon Electrode

Yuanyuan Zhang,Qi Cao,Fengmei Zhu,Hedan Xu,Yang Zhang,Wenfeng Xu,Xiaoling Liao
DOI: https://doi.org/10.20964/2020.09.62
IF: 1.541
2020-01-01
International Journal of Electrochemical Science
Abstract:An enzyme-free hydrogen peroxide (H2O2) electrochemical sensor was constructed through building carbon nanotubes (CNTs), reduced graphene oxide (RGO) and platinum nanoparticles (Pt NPs) nanocomposite with a 3D structure modified glassy carbon electrode (GCE). Scanning electron microscopy (SEM) and energy -dispersive X-ray spectroscopy (EDS) analysis were used to characterize the morphology and element distribution of the modified electrodes RGO/CNTs-Pt/GCE. The electrocatalytic reduction performance of RGO/CNTs-Pt on H2O2 was investigated by cyclic voltammetry (CV) and chronoamperometry (i-t). Because of the synergistic effect of enhanced conductivity and catalytic ability of the carbon based nanocomposites (RGO/CNTs) and Pt NPs implanted 3D interfaces, the sensor shows excellent electrocatalytic property on H202 reduction. Under the optimized conditions, a high sensitivity (347 +/- 5 mu A mM(-1) cm(-2), n = 3) and a linear range (0.0003 - 0.018 mM and 0.01 - 4.0 mM) with a low detection limit (similar to 0.31 mu M) operated at the working voltage of -0.2 V (vs. Ag/AgCl) are achieved by RGO/CNTs-Pt/GCE. In addition, the sensor is resistant to interference and maintains long-term stability and has been successfully used to detect the spiked H2O2 in milk samples, which has important significance for monitoring the abuse of H2O2 in milk.
What problem does this paper attempt to address?