Simulation and Calibration of the Response Function of Multi-Sphere Neutron Spectrometer

Jiang Xiaofei,Chen Chao,Cao Jing,Cao Hongrui,Li Shiping,Song Xianying,Yin Zejie
DOI: https://doi.org/10.13538/j.1001-8042/nst.2013.06.010
2014-01-01
Abstract:In order to realize the on-line real-time measurement of neutron spectrum of ITER fusion, this paper presents a multi-sphere spectrometer system which consists of eight thermal neutron detectors, namely SP9 He-3 proportional counter, embedded in eight different diameter polyethylene spheres. The response function of eight polyethylene spheres of multi-sphere neutron spectrometer was calculated after the simulation of the neutron transport processes in multi-sphere spectrometer by adopting software Geant4. The peak of the response function is in the low energy region for smaller diameter polyethylene sphere. As the polyethylene sphere diameter increased, the peak of the response function moves to the high energy region. The experimental calibration adopts Am-241-Be neutron source. The relative error between normalized data of experiment 4 pi solid angle counts and normalized data of simulated detection efficiency of 4in to 8in polyethylene sphere is from 1.152% to 12.222%.The experimental results verify the response function of the simulation. All these results provide a theoretical and experimental basis for solving the on-line real-time neutron spectrum of ITER fusion.