316L Stainless Steel Manufactured by Selective Laser Melting and Its Biocompatibility with or without Hydroxyapatite Coating

Jiapeng Luo,Xiao Jia,Ruinan Gu,Peng Zhou,Yongjiang Huang,Jianfei Sun,Ming Yan
DOI: https://doi.org/10.3390/met8070548
IF: 2.695
2018-01-01
Metals
Abstract:To fabricate metallic 316L/HA (hydroxyapatite) materials which meet the requirements of an implant's mechanical properties and bioactivity for its function as human bone replacement, selective laser melting (SLM) has been employed in this study to prepare a 316L stainless steel matrix, which was subsequently covered with a hydroxyapatite (HA) coating using the sol-gel method. High density (98.9%) as-printed parts were prepared using a laser power of 230 W and a scanning speed of 800 mm/s. Austenite and residual acicular ferrite existed in the microstructure of the as-printed 316L stainless steel, and the sub-grain was uniform, whose primary dendrite spacing was around 0.35 m. The as-printed 316L stainless steel showed the highest Vickers hardness, elastic modulus, and tensile strength at similar to(similar to means about; same applies below unless stated otherwise) 247 HV, similar to 214.2 GPa, and similar to 730 MPa, respectively. The elongation corresponding to the highest tensile strength was similar to 38.8%. The 316L/HA structure, measured by the Relative Growth Rate (RGR) value, exhibited no cell cytotoxicity, and presented better biocompatibility than the uncoated as-printed and as-cast 316L samples.
What problem does this paper attempt to address?