Improved Chemical Exchange Saturation Transfer Imaging with Real‐time Frequency Drift Correction

Ruibin Liu,Hongxi Zhang,Weiming Niu,Can Lai,Qiuping Ding,Weibo Chen,Sayuan Liang,Jinyuan Zhou,Dan Wu,Yi Zhang
DOI: https://doi.org/10.1002/mrm.27663
2019-01-01
Magnetic Resonance in Medicine
Abstract:Purpose: To investigate the effects of frequency drift on chemical exchange saturation transfer (CEST) imaging at 3T, and to propose a new sequence for correcting artifacts attributed to B-0 drift in real time. Theory and Methods: A frequency-stabilized CEST (FS-CEST) imaging sequence was proposed by adding a frequency stabilization module to the conventional non-frequency-stabilized CEST (NFS-CFST) sequence, which consisted of a small tip angle radiofrequency excitation pulse and readout of three non-phase-encoded k-space lines. Experiments were performed on an egg white phantom and 26 human subjects on a heavy-duty clinical scanner, in order to compare the difference of FS-CEST and NFS-CEST sequences for generating the z-spectrum, magnetization transfer ratio asymmetry (MTRasym) spectrum, and amide proton transfer weighted (APTw) image. Results: The B-0 drift in CEST imaging, if not corrected, would cause APTw images and MTRasym spectra from both the phantom and volunteers w be either significantly higher or lower than the true values, depending on the status of the scanner. The FS-CEST sequence generated substantially more stable MTR(asym )spectra and APTw images than the conventional NFS-CEST sequence. Quantitatively, the compartmental-average APTw signals (mean +/- standard deviation) from frontal white matter regions of all 26 human subjects were -0.32% +/- 2.32% for the NFS-CEST sequence and -0.14% +/- 0.37% for the FS-CEST sequence. Conclusions: The proposed FS-CEST sequence provides an effective approach for B-0 drift correction without additional scan time and should be adopted on heavy-duty MRI scanners.
What problem does this paper attempt to address?