Monte Carlo Calculations for Reporting Patient Organ Doses from Interventional Radiology

Wanli Huo,Mang Feng,Yifei Pit,Zhi Chen,Yiming Gao,X. George Xu
DOI: https://doi.org/10.1051/epjconf/201715304016
2017-01-01
EPJ Web of Conferences
Abstract:This paper describes a project to generate organ dose data for the purposes of extending VirtualDose software from CT imaging to interventional radiology (IR) applications. A library of 23 mesh-based anthropometric patient phantoms were involved in Monte Carlo simulations for database calculations. Organ doses and effective doses of IR procedures with specific beam projection, filed of view (FOV) and beam quality for all parts of body were obtained. Comparing organ doses for different beam qualities, beam projections, patients’ ages and patient’s body mass indexes (BMIs) which generated by VirtualDose-IR, significant discrepancies were observed. For relatively long time exposure, IR doses depend on beam quality, beam direction and patient size. Therefore, VirtualDose-IR, which is based on the latest anatomically realistic patient phantoms, can generate accurate doses for IR treatment. It is suitable to apply this software in clinical IR dose management as an effective tool to estimate patient doses and optimize IR treatment plans.
What problem does this paper attempt to address?