Dual-dircrtion Excitation Dissociation Technology in Digital Ion Ttrap

Xu Fu-Xing,Chen Xin,Chen Zhi-Yang,Zhou Ming-Fei,Ding Chuan-Fan
DOI: https://doi.org/10.19756/j.issn.0253-3820.191016
IF: 1.193
2019-01-01
Chinese Journal of Analytical Chemistry
Abstract:Tandem mass spectrometric analysis is one primary function of ion trap mass analyzer. In linear ion trap, this function is conducted by collision induced dissociation (CID), which is realized via applying dipolar excitation voltage on one couple of electrodes, so the CID method is also known as unidirectional ion excitation. In this work, bidirectional resonance excitation method was introduced. By applying dipolar excitation voltage on two couples of electrodes, the ion excitation energy and associated dissociation efficiency of this method were increased. The simulation and experimental results indicated that based on this method, higher ion dissociation efficiency was obtained with different q values and excitation frequencies. Besides, more small fragments could be detected in tandem mass analysis of leucine enkephalin via bidirectional resonance excitation with q= 0.352, V= 0. 5 V0-p, omega = 107.93 kHz and t = 5 ms, suggesting that the method could effectively decrease low mass cut-off effect and strength ability of ion trap on tandem mass analysis.
What problem does this paper attempt to address?