Rapid Analysis of Bisphenol A in Food Packaging Materials by Desorption Corona Beam Ionization Tandem Mass Spectrometry

Wang Yan-Jiao,Li Xiang,Sun Wen-Jian,Mu Hui,Zhang Xiao-Qiang,Ma Hong-Ying,Xu Ting-Ting,Gao Chao
DOI: https://doi.org/10.3724/SP.J.1096.2014.30567
IF: 1.193
2014-01-01
Chinese Journal of Analytical Chemistry
Abstract:A novel method, based on desorption corona beam ionization (DCBI) tandem mass spectrometry (MS), was developed for the direct detection of bisphenol A in food packaging materials with minimal sample pretreatment. We optimized DCBI-MS conditions (i.e. extraction solvent and ionization source) using representative bisphenol A standards. The reagent solvent was methanol at high voltage (-2.0 kV, negative ion detection mode). The temperature of the heated capillary was optimized to be 225 degrees C. The flow rate of He was 1.0 L/min. The limit of detection (LOD) of these amino acids was < 0.1 mg/L, and the linear dynamic range was 1-100 mg/L. The bisphenol A concentration in food packaging material samples was measured by the method, and the average time for a single sample analysis was less than 30 s. The recovery rates for different concentrations ranged from 94% to 125%, and the relative standard deviation (RSD) was less than 17%. This study showed that DCBI-MS was a powerful tool for the rapid, sensitive, and quantitative detection of bisphenol A amino acids in complex meterials samples.
What problem does this paper attempt to address?