Determination of 4 Sterols in Atmospheric Particles by Ultrasonic Extraction Coupled with High Performance Liquid Chromatography-Tandem Mass Spectrometry

Zhao Qi-Yue,Zhao Hong-Shuai,Liu Bao-Xian,Wang Yan,Luo Fang
DOI: https://doi.org/10.11895/j.issn.0253.3820.150783
IF: 1.193
2016-01-01
Chinese Journal of Analytical Chemistry
Abstract:A rapid analytical method was developed for the determination of 4 sterols in atmospheric particles by ultrasonic extraction coupled with high performance liquid chromatography-tandem mass spectrometry (HPLC-MS/MS). After extraction by methanol and enrichment, sterols were separated on an Atlantis C-18 column (100 mmx2.1 mm, 3 mu m) with acetonitrile and water served as mobile phase in gradient elution, then determined by HPLC-MS/MS using atmospheric pressure chemical positive ionization (APCI(+)) and multiple reaction monitoring (MRM). Parameters of chromatography and mass spectrometry, as well as conditions of ultrasonic extraction were optimized. The developed method provided good linearity (R > 0. 99) and high recoveries (80.3%-97.7%) for 4 sterols. Low detection limits were obtained as 0.015 ng/m(3), with RSD less than 15% (n = 6). Comparing with GC/MS, this method was more rapid, sensitive and economical because the derivation was eliminated from the procedure, thus saving a lot of resources. Samples of atmospheric particles were detected using this method, and the results indicated that this method could meet the requirements for quantitative analysis of sterols in atmosphere.
What problem does this paper attempt to address?