Determination of Polychlorinated Naphthalenes in Soil and Sediment Samples by Accelerated Solvent Extraction-Multilayer Silica Gel Column Cleanup Coupled with Gas Chromatography-Triple Quadrupole Mass Spectrometry

Zhang Li-Fei,Zhang Xiu-Lan,Zhang Hui,Li Ling-Ling,Zhang Lin-Li,Dong Liang
IF: 1.193
2014-01-01
Chinese Journal of Analytical Chemistry
Abstract:A method for determining polychlorinated naphthalenes (PCNs) in soil and sediment samples has been developed by using accelerated solvent extraction-multilayer silica gel column cleanup coupled with gas chromatography. triple quadrupole mass spectrometry (GC-QqQ-MS/MS). The cleanup condition based on commercial multilayer silica gel column was optimized. 150 mL hexane was finalized and used as eluent. D7-mono chlorinated naphthalene (CN) (#2) was used as internal standard for monoCN and diCN. Main parameters of GC and MS/MS were optimized as well to improve the conditions of separation and resolution of target compounds. Baseline separation of hexaCN (#66) and hexaCN (#67) was accomplished by GC. QqQ-MS/MS using a Rt-beta DEXcst capillary column in multiple reaction monitoring mode. The relative standard deviations of the average relative response factors for the 18 PCN congeners were below 18.5% in the range of 1.0-240 mu g/L. The limits of detection and limits of quantitation of PCNs were in the range of 0.014-0.858 mu g/L and 0.048-2.862 mu g/L, respectively. The overall performance of the 30 meters Rt-beta DEXcst capillary column was better than the traditional 60 meters DB-5MS column. However, it was time. consuming to use the former column. The present method was applied for the determination of PCNs in soil and sediment samples from the Southern Jiangsu Province. The recoveries of the seven labeled PCNs were higher than 28%, except for monoCN (about 6%) in the real samples. The results showed that lower chlorinated naphthalenes were the predominant congeners in the test samples.
What problem does this paper attempt to address?