Simultaneous Detection of Mercury, Lead and Strontium Ions Based on Conformational Conversion Sensor

Zhou Yuting,Tang Yujiao,Shao Shuang,Dai Shiyan,Cheng Guifang,He Pingang,Fang Yuzhi
DOI: https://doi.org/10.7503/cjcu20180868
2019-01-01
Abstract:A high. sensitive and specific electrochemical biosensor was designed for simultaneous detection of mercury ion(Hg2+), lead ion(Pb2+) and strontium ion(Sr2+) by oligonucleotide conformational transformation. The biosensor was based on porous anodic alumina(PAA) modified with two oligonucleotide aptamers which could capture Hg2+, Pb2+ and Sr2+, respectively. The aptamers can block [Fe(CN)(6)](3-/4-) ion conduction in the nanochannels of PAA. In the presence of Hg2+, Pb2+ and Sr2+, the aptamers could bend and wrap around those metal ions to make the nanochannel unobstructed, and then, the redox current of [Fe(CN)(6)](3-/4-) could be increased. When the biosensor eluted with beta-mercaptoethanol(beta ME), 1,4,7,10-tetraazacyclododecane-1,4,7,10-tetraacetic acid(DOTA) and kryptofix 222(KP) successively, the aptamers stretched and released Hg2+, Pb2+ and Sr2+ in sequence, leading to the re-blockage of nanochannel with the decrease of the redox current. The calibration curves of Hg2+, Pb2+ and Sr2+ were obtained in the range of 0.05-2.50 nmol/L, and the detection limits were 0.013, 0.017 and 0.022 nmol/L, respectively. Further. more, the biosensor is easily constructed, stable and reusable. It is hopeful to apply in the daily monitoring of harmful metal ions in drinking water or environmental water.
What problem does this paper attempt to address?