Real-time characterization and quantitative detection of cytochrome c based on SAMs with electrochemical quartz crystal microbalance

JH Li,JB Hu,QL Li
DOI: https://doi.org/10.3321/j.issn:0251-0790.2005.06.008
2005-01-01
Abstract:Electrochemical quartz crystal microbalance (EQCM) which is capable of sensitive measurement for mass and electrochemical signals, was employed to study cytochrome c. Self-assembled monolays (SAMs) with carboxy groups were first coated on the gold electrode of a quartz crystal resonator by adsorption of omega-mercaptbundecanoic acid(MUA), cytochrome c was then covalently attached to the surface via N-ethyl-N'-[3-(dimethyl)aminopropyl] carbodiimide hydrochloride (EDC) and N-hydroxysuccinimide (NHS). The processes for self-assembly of MUA and immobilization of cytochrome c were both real-time characterized. And the surface coverage densities of MUA and cytochrome c were estimated to be 3.73 X 10(-9) and 7.31 X 10(-11) mol/cm(2), respectively. The amount of cytochrome c was found to be proportional to the frequency changes between 0.03 mu mol/L and 3.0 mu mol/L, with a regression equation Delta f (Hz) = 50.19 c(Cyt C) + 67.07 (r(2) = 0.998), and the limit of detection of 1.19 X 10(-9) mol/L was also attained.
What problem does this paper attempt to address?